Design of bluetooth lung function instrument |
LIANG Rui1 YAO Kainan2 DENG Yingqiang3 CHEN Jianmin3 |
1.Department of Pediatrics, the First Affiliated Hospital of Guangzhou University of Chinese Medicine, Guangdong Province, Guangzhou 510405, China; 2.Department of Pulmonary Diseases, Shenzhen Bao′an Hospital of Traditional Chinese Medicine, Guangdong Province, Shenzhen 518101, China;
3.Department of Research and Development, Guangzhou Changhu Medical Devices Co. Ltd, Guangdong Province, Guangzhou 510000, China |
|
|
Abstract Objective To design and develop a bluetooth portable device which can detect common lung function parameters. Methods The data of high precision varistor were collected and processed by single chip microcomputer. And the test data were calculated by calibration parameters and linear interpolation. Standard breathing simulator error and actual measurement error for testing bluetooth lung function meter. Results The bluetooth lung function instrument was designed successfully, which could monitor forced vital capacity (FVC), peak expiratory flow, forced expiratory volume in one second (FEV1) and FEV1/FVC. The requirement was met by the bluetooth pulmonary function. Compared with the yeger pulmonary function tester, the measuring error by the bluetooth pulmonary function tester of six subjects were all within 5%. Conclusion The device is small and convenient, and can meet the application requirements of clinical and home tests.
|
|
|
|
|
[1] 陈君石,黄建始.健康管理师[M].北京:中国协和医科大学出版社,2007:11-18.
[2] 郭清.智能健康管理[J].健康研究,2011,31(2):81-85.
[3] World Health Organization. A universal truth:no health without a workforce [EB/OL]. [2013-02-24]. http://www.who.int/workforcealliance/knowledge/resources/hrhreport2013/en/index.html.
[4] 汤胜蓝,Ehiri J,龙倩.慢性非传染性疾病:中国卫生领域被忽视的最大挑战[J].中国卫生政策研究,2013,6(10):6-11.
[5] 席晶晶,夏小亮,司向,等.2011年全国疾控机构慢性病预防控制人力资源配置的现状分析[J].中国慢性病预防与控制,2014,22(3):287-289.
[6] Whitehead L,Seaton P. The effectiveness of self-management mobilephone andtablet apps in long-term condition management:a systematic review [J]. J Med Internet Res,2016,18(5):e97.
[7] Adler-Milstein J,Kvedar J,Bates DW. Telehealth among US hospitals:several factors,including state reimbursement and licensure policies,influence adoption [J]. Health Aff(Milwood),2014,33(2):207-215.
[8] Masoli M,Fabian D,Holt S,et al. The global burden of asthma:executivesummary of the GINA Dissemination Committee report [J]. Allergy,2004,59(5):469-478.
[9] Pocket Guide for Asthma Management and Prevention. Global Initiative for Asthma(GINA)[J]. 2011, Available from:http://www.ginasthma.org/.
[10] 刘传合,陈育智.儿童哮喘流行病学及防治现状分析[J].中国实用儿科杂志,2013,28(11):809-811.
[11] 全国儿科哮喘协作组,中国疾病预防控制中心环境与健康相关产品安全所.第三次中国城市儿童哮喘流行病学调查[J].中华儿科杂志,2013,51(10):729-735.
[12] 鲍一笑.儿童支气管哮喘诊断与防治指南(2016年版)[C].中国中西医结合学会儿科专业委员会.第二十次全国儿科中西医结合学术会议资料汇编.中国中西医结合学会儿科专业委员会:中国中西医结合学会,2016:57-71,56.
[13] 何孝晶.心理护理联合健康教育在肺功能检测中的应用效果观察[J].心理月刊,2019,14(16):62.
[14] 阙瑞英.对行肺功能检查的儿童实施优质护理的效果探析[J].当代医药论丛,2017,15(11):166-168.
[15] 李树民,张敏,王朋,等.小鼠无创性肺功能的测定及其意义[J].中国实验动物学报,2018,26(5):548-553.
[16] 柏玲,宋新,李晨霞,等.峰流速仪在支气管哮喘的临床研究现状[J].世界最新医学信息文摘,2019,19(74):41,48.
[17] 黄燕,蒋芸,孙纪光,等.新型运动心电图检测系统及初步临床应用[J].北京生物医学工程,2018,37(3):303-310.
[18] 何攀,冯伟.婴幼儿哮喘控制评估工具的临床应用进展[J].西南军医,2019,21(2):161-165.
[19] 王开梅,胡祥英,郑光强,等.外周血ADAM8、IRF1表达水平与支气管哮喘患儿肺功能的相关性研究[J].中国医药导报,2019,16(24):95-98,106.
[20] 冯艳华.哮喘儿童家庭管理的研究现状[J].全科护理,2017,15(10):1185-1188.
[21] 费霞,张雪,张旻,等.呼气峰流速仪在哮喘诊治中的应用价值[J].上海医学,2017,40(2):126-128.
[22] 包春燕.电子峰流速仪在支气管哮喘住院患者中的应用及护理[J].影像研究与医学应用,2018,2(9):223-225.
[23] 吴寿萍,马璐,吴绮,等.电子峰流速仪在支气管哮喘住院患者中的应用及护理效果观察[J].名医,2019(1):162. |
|
|
|