Study on the difference of Keratron corneal ultrasound thickness gauge and Pentacam in the measurement of central corneal thickness in patients with refractive error in Inner Mongolia Region |
WANG Zhaoge ZHAO Haixia GUAN Wenying GE Ruichun WANG Ruifang |
Myopia Laser Treatment Centre, the Affiliated Hospital of Inner Mongolia Medical University, Inner Mongolia Autonomous Region, Hohhot 010050, China |
|
|
Abstract Objective To compare the difference in central corneal thickness measurement of patients with refractive error in Inner Mongolia Region performed with Keratron corneal ultrasound thickness gauge and Pentacam, and provide reference for the analysis of corneal biomechanical parameters and basis for the safety of excimer laser surgery. Methods From March 2015 to November 2016, 140 patients with refractive error were randomly selected from the Myopia Laser Treatment Center of the Affiliated Hospital of Inner Mongolia Medical University, including 70 Mongolian (137 eyes) and 70 Han (140 eyes) patients. Pentacam and Keratron corneal ultrasound thickness gauge were used to measure the central corneal thickness of Mongolian and Han patients with refractive error in Inner Mongolia Region, and the data of the measurement results were analyzed. Results The central corneal thickness of Mongolians patients was (549.02±33.55) μm, that of Han patients was (575.92±37.19) μm measured with Keratron, the central corneal thickness of Mongolians patients was (538.34±32.61) μm, that of Han patients was (557.75±31.99) μm measured with Pentacam, there were high statistically significant differences in central corneal thickness for the different nationalities measured with same equipment and for the same nationality measured with different equipment (P < 0.01). Conclusion The thickness of the central cornea measured by Pentacam was thinner than that of Keratron corneal ultrasound thickness gauge, and the thickness of the central cornea of the Mongolian patients in the study objects is thinner than that of the Han patients.
|
|
|
|
|
[1] Park SH,Choi SK,Lee D,et al. Corneal thickness measurement using Orbscan,Pentacam,Galilei,and ultrasound in normal and post-femtosecond laser in situ keratomileusis eyes [J]. Cornea,2012,31(9):978-982.
[2] Randleman JB,Lynn MJ,Perez-Straziota CE,et al. Comparison of central and peripheral corneal thickness measurements with scanning-slit,Scheimpflug and Fourier-domain ocular coherence tomography [J]. Br J Ophthalmol,2015,99(9):1176-1181.
[3] Rio-Cristobal A,Martin R. Corneal assessment technologies:current status [J]. Surv Ophthalmol,2014,59(6):599-614.
[4] Yap TE,Archer TJ,Gobbe M,et al. Comparison of central corneal thickness between fourier-domain oct,very high-frequency digital ultrasound,and Scheimpflug imaging systems [J]. J Refract Surg,2016,32(2):110-116.
[5] Maloca PM,Studer HP,Ambrósio R,et al. Interdevice variability of central corneal thickness measurement [J]. PLoS One,2018,13(9):e0203884.
[6] Feizi S,Jafarinasab MR,Karimian F,et al. Central and peripheral corneal thickness measurement in normal and keratoconic eyes using three corneal pachymeters [J]. J Ophthalmic Vis Res,2014,9(3) :296-304.
[7] Yildirim Y,Olcucu O,Agca A,et al. Topographic and biomechanical evaluation of corneas in patients with ocular rosacea [J]. Cornea,2015,34(3):313-317.
[8] Beutelspacher SC,Serbecic N,Scheuerle AF. Assessment of central corneal thickness using OCT,ultrasound,optical low coherence reflectometry and Scheimpflug pachymetry [J]. Eur J Ophthalmol,2011,21(2):132-137.
[9] 王霁雪,吴荒,杨隆艳,等. Orbscan-Ⅱ、Pentacam及超声测厚仪角膜厚度测量值的比较[J].中华眼视光学与视觉科学杂志,2010,12(1):64-67.
[10] 王淑荣,肖鹏柁,刘鑫,等.角膜厚度测量影响因素研究进展[J].中国实验诊断学,2015,19(10):1805-1808.
[11] 李坤,陆蓓,崇晓霞,等.麻醉剂和泪膜对健康人角膜厚度的影响[J].眼视光学杂志,2004,6(4):244-246.
[12] 王华玲,赖江龙.不同表面麻醉剂在眼科A超检查中对角膜上皮的损伤研究[J].国际眼科杂志,2014,20(11):29-34.
[13] Niimi J,Tan B,Chang J,et al. Diurnal Pattern of Tear Osmolarity and Its Relationship to Corneal Thickness and Deswelling [J]. Cornea,2013,32(10):1305-1310.
[14] Zhao HX,Zhang L,Guan WY,et al. Difference in normal corneal thickness and curvature between Mongolian and Han nationalities [J]. Int J Ophthalmol,2015,8 (2):399-402.
[15] Naderan M,Shoar S,Naderan M,et al. Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography [J]. Int J Ophthalmol,2015,8(2):275-280.
[16] Wongchaisuwat N,Metheetrairat A,Chonpimai P,et al. Comparison of central corneal thickness measurements in corneal edema using ultrasound pachymetry,Visante anterior-segment optical coherence tomography,Cirrus optical coherence tomography,and Pentacam Scheimpflug camera tomography [J]. Clin Ophthalmol,2018,25(12):1865-1873.
[17] Nassiri N,Sheibani K,Safi S,et al. Central corneal thickness in highly myopic eyes:inter-device agreement of ultrasonic pachymetry,pentacamand orbscan Ⅱ before and after photorefractive keratectomy [J]. J Ophthalmic Vis Res,2014,9(1):14-21.
[18] Nemeth G,Hassan Z,Szalai E,et al. Anterior segment parameters measured with 2 optical devices compared to ultrasonic data [J]. Eur J Ophthalmol,2013,23(2):177-182.
[19] Faria-Correia F,Ramos I,Valbon B,et al. Scheimpflug-based tomography and biomechanical assessment in pressure-induced stromal keratopathy [J]. J Refract Surg,2013, 29(5):356-358.
[20] Sedaghat MR,Daneshvar R,Kargozar A,et al. Comparison of central corneal thickness measurement usingultrasonic pachymetry,rotating Scheimpflug camera,and scanning-slit topography [J]. Am J Ophthalmol,2010, 150(6):780-789.
[21] 郝更生,曾莉,李玉茹,等.Pentacam系统与A超角膜测厚仪测量中央角膜厚度的一致性和重复性[J].中华眼科杂志,2011,47(2):142-145. |
|
|
|