The clinical value of microfluidic loop-mediated isothermal amplification detection in the diagnosis of Klebsiella pneumoniae infection in sepsis patients |
SU Hua1 CHEN Chen2 JIANG Li′na3 JIAQian1 XU Yuhuan1 LI Xiaofeng1 LIU Yunning1 LI Jie1 |
1.Institute of Laboratory Medicine, the First Affiliated Hospital of Hebei North University, Hebei Province, Zhangjiakou 075000, China;
2.Department of Intensive Care Unit, the First Affiliated Hospital of Hebei North University, Hebei Province, Zhangjiakou 075000, China;
3.Hebei North University, Hebei Province, Zhangjiakou 075000, China |
|
|
Abstract Objective To explore the clinical value of microfluidic loop-mediated isothermal amplification detection in the diagnosis of Klebsiella pneumoniae infection in sepsis patients. Methods From June 2015 to May 2018, 152 sepsis patients who were from the department of ICU of the First Affiliated Hospital of Hebei North University were collected, 99 cases of sputum specimens samples were collected by natural cough method, 53 cases were collected by bronchoscopy. All these samples were simultaneously detected by traditional pathogen culture method and microfluidic loop-mediated isothermal amplification method. According to the traditional bacterial culture result, all cases were divided into two group, 42 cases were Klebsiella pneumoniae infection group and 110 cases were non Klebsiella pneumoniae infection group. Results The positive rate of Klebsiella pneumoniae detected by bacterial culture was 27.6%(42/152), and by microfluidic loop-mediated isothermal amplification detection was 38.8%(59/152), which was higher than the former, the difference was statistically significant(P < 0.01). The method of microfluidics loop-mediated isothermal amplification detected 37 positive cases from Klebsiella pneumoniae infection group, including 21 cases natural cough sample and 16 cases bronchoscopy sample, Non Klebsiella pneumoniae infection group was detected by the method of microfluidics loop-mediated isothermal amplification had 22 cases positive samples,including 15 cases natural cough samples and 7 cases bronchoscopy samples. The area under the ROC curve of the bronchoscopy suction method and natural cough collection method was 0.836 and 0.847. The Jordan index of former (0.694) was high er than the latter (0.672). The Kappa value was 0.604 > 0.600, P < 0.05, indicating a high consistency between the two methods. Conclusion Microfluidic loop-mediated isothermal amplification has the advantages of rapid and sensitive detection of Klebsiella pneumoniae in sepsis patients, which is highly consistent with the traditional bacterial culture method and can provide better clinical diagnosis and treatment basis.
|
|
|
|
|
[1] 林文慧,邹秉杰,宋沁馨,等.多重环介导等温扩增技术研究进展[J].遗传,2015,37(9):899-910.
[2] Zhang C,Yao Y,Zhu JL,et al. Establishment and application of a real-time loop-mediated isothermal amplification system for the detection of CYP2C19 polymorphisms [J]. Sci Rep,2016,1(6):26533.
[3] 范菲楠,聂署萍,萧晓友,等.微流控技术检测宫颈脱落细胞HPV的临床应用[J].临床检验杂志:电子版,2018, 7(3):375-377.
[4] 黄世光,靳翔宇,林荣赞,等.微流控芯片核酸分析系统及其精准医学应用[J].中国激光,2018,45(3):124-131.
[5] 王云超,牛海涛,李凯,等.微流控芯片特异性筛选膀胱癌循环肿瘤细胞的研究[J].中华泌尿外科杂志,2016,37(12):936-939.
[6] 邹丽辉,李英,肖飞.微流控技术在临床检验中的应用[J].中华检验医学杂志,2018,41(9):631-633.
[7] Loo JF,Kwok HC,Leung CC,et al. Sample-to-answer on molecular diagnosis of bacterial infection using integrated lab—on—a—disc [J]. Biosens Bioelectron,2017,15(93):212-219.
[8] 沈韧,万谅,贾艳伟.微流控技术在临床检测中的应用[J].分子诊断与治疗杂志,2018,10(5):289-294.
[9] 徐欢,陈鸣.微流控生物传感器及其分子诊断应用[J].中华检验医学杂志,2018,41(9):634-637.
[10] Zhu Z,Tang Y,Jiang YS,et al. Strand-Exchange Nucleic Acid Circuitry with Enhanced Thermo-and Structure- Buffering Abilities Turns Gene Diagnostics Ultra-Reliable and Environmental Compatible [J]. Sci Rep,2016,4(6):36 605.
[11] 舒博文,林冬果,雷秀霞,等.微流控体外诊断技术应对临床检验医学的挑战[J].中华检验医学杂志,2018,41(9):696-699.
[12] Seyrig G,Stedtfeld RD,Tourlousse DM,et al. Selection of fluorescent DNA dyes for real-time LAMP with portable and simple optics [J]. Microbiol Methods,2015,11(119):223-227.
[13] Chaumpluk P,Plubcharoensook P,Prasongsuk S. Rapid detection of aflatoxigenic Aspergillus sp.in herbal specimens by a simple,bendable,paper-based lab-on-a-chip [J]. Biotechnol,2016,11(6):768-779.
[14] 李松,何晓光,马可泽.环介导恒温扩增芯片法在机械通气新生儿下呼吸道病原菌检测中应用[J].广东医科大学学报,2018,36(2):169-172.
[15] 钱婧,朱春梅,曹玲,呼吸道病原菌基因检测法诊断儿童肺炎链球菌感染的临床价值[J].中华全科医学,2018, 16(9):1512-1520.
[16] Badolo A,Bando H,Traoré A,et al. Detection of G119S ace-1(R)mutation in field-collected Anopheles gambiae mosquitoes using allele-specific loop-mediatedisothermal amplification(AS-LAMP) method [J]. Malar J,2015, 14(1):477.
[17] Hsieh K,Ferguson BS,Eisenstein M,et al. Integrated electrochemical microsystems for genetic detection of pathogens at the point of care [J]. Acc Chem Res,2015, 48(4):911-920.
[18] 杜芳玲,王婷婷.环介导等温扩增(LAMP)技术的研究进展与展望[J].实验与检验医学,2018,36(4):509-512, 565.
[19] 张亚平,郑志高,简保磊,等.新型环介导等温扩增技术开发与应用研究进展[J].分子诊断与治疗杂志,2018, 10(2):138-144.
[20] 温福利.兔斯氏艾美耳球虫环介导等温扩增技术检测方法的建立[J].实验动物与比较医学,2018,38(3):176-181. |
|
|
|