Comparison of application effects of different systems in continuous invasive blood pressure monitoring in elderly patients with septic shock#br# |
LIU Guangdi1,2 TANG Li1,2 LIU Yiwen2,3▲ LI Qian2 |
1.Department of Respiratory and Critical Care Medicine, West China Hospital, Sichuan University, Sichuan Province, Chengdu 610041, China;
2.West China College of Nursing, Sichuan University, Sichuan Province, Chengdu 610041, China;
3.Department of Critical Care Medicine, West China Hospital, Sichuan University, Sichuan Province, Chengdu 610041, China |
|
|
Abstract Objective To explore the effect of different systems in the protection of access for continuous invasive blood pressure monitoring in elderly patients with septic shock. Methods A total of 80 elderly patients with septic shock who were placed with arterial pressure monitoring catheters in West China Hospital of Sichuan University from May 2020 to May 2021 were selected and divided into experimental group and control group by random number table method, with 40 cases in each group. The experimental group was given a total closed system to monitor the arterial pressure for access protection, while the control group was given the general system. The postoperative catheter retention time, catheter complications, stability, and comfort of the arterial pressure measurement system were compared between the two groups. Results The catheter indwelling time in the experimental group was longer than that in the control group, while the time for a single collection of blood gas analysis specimens in the experimental group was shorter than that in the control group, and the difference was statistically significant (P < 0.05). The bleeding rate, swelling rate, tube blocking rate, and complication rate in the experimental group were lower than those in the control group, and the differences were statistically significant (P < 0.05). There was no statistically significant difference in the number of abnormal numerical values and the number of abnormal waveforms between the two groups (P > 0.05). The comfort level of the experimental group was higher than that of the control group, and the difference was statistically significant (P < 0.05). Conclusion The fully closed system has the same effect as the general system in maintaining the stability of the arterial pressure measurement system, which can reduce the occurrence of complications and improve the patient’s comfort.
|
|
|
|
|
[1] 罗运山,罗君,刘易林,等.脓毒性休克预后因素分析[J].重庆医学,2017,46(19):129-131.
[2] 曹钰,柴艳芬,邓颖,等.中国脓毒症/脓毒性休克急诊治疗指南(2018)[J].临床急诊杂志,2018,19(9):567-588.
[3] 钱嘉俊.脓毒症相关性脑病一例并文献复习[D].广州:广州医科大学,2020.
[4] 吴鹏.脓毒性休克早期液体复苏后血流动力学指标变化趋势分析及临床意义研究[D].青岛:青岛大学,2019.
[5] 黄慧钟,许云香.有创血压监测在进展为脓毒性休克患者中的应用分析[J].临床急诊杂志,2018,19(6):410-412.
[6] 刘爽,李恩有.动脉血压监测的临床应用[J].医学综述,2017,23(9):1771-1774,1781.
[7] Leslie R,AGouldson S,Habib N,et al. Management of Arterial Lines land Blood Sampling in Intensive Care: A Threat to Patient Safety [J]. Anaesthesia,2013,68(11):1114-1119.
[8] 罗玉珍,汪海芹,高明珠,等.不同部位动脉置管监测有创血压ICU危重患者中应用效果[J].临床护理志,2016, 15(6):27-29.
[9] 王轶,韩柳,袁翠,等.成人ICU患者外周动脉导管留置与维护的最佳证据总结[J].中华护理杂志,2020,55(4):600-606.
[10] 郭汉画,陈名桂,孔丽丽,等.重症患者动脉测压导管最佳更换策略的循证实践[J].护理学报,2021,28(2):37-41.
[11] 杨晓玲,冷志兵,江湖,等.有创动脉血压监测中生理盐水和肝素生理盐水冲管效果比较的系统评价[J].中国护理管理,2018,18(10):1332-1339.
[12] 任玉娇,朱丽萍.巧用充气式压迫器制作简易Allen试验用具[J].中华现代护理杂志,2014,20(36):4660.
[13] 舒雅娟.有创动脉血压监测在ICU危重症患者中的应用及护理[J].中西医结合护理:中英文,2017,3(3):95-97.
[14] 曹岚,张丽娜,艾关林,等.超声引导下桡动脉穿刺置管在重症患者中的应用[J].中南大学学报:医学版,2018, 43(4):447-451.
[15] Arthur CT,Gilles C,Michael FO,et al. Indications,interpretation,and techniques for arterial catheterization for invasive monitoring[EB/OL].(2019-07-24)[2019-09-04].
[16] 罗彩远,黎艳.正压输液接头运用联合集束化护理干预在重症患者有创血压监测中的效果评价[J].护士进修杂志,2019,34(7):648-650.
[17] 胥小芳,孙红,李春燕,等.动脉血气分析临床操作实践标准要点解读[J].中国护理管理,2017,17(9):1158-1161.
[18] 杨肖蓉,刘建波.静脉/动脉血液管理保护通路(Venous and Arterial Monitor Protection,VAMP)在ICU危重病人中的应用[J].世界最新医学信息文摘,2013,13(15):477.
[19] 郑菁,陈鑫鑫,林娜.生理盐水持续冲洗动脉测压管的效果评价[J].福建医药杂志,2015,37(2):159-160.
[20] 刘爽,李恩有.动脉血压监测的临床应用[J].医学综述,2017,23(9):1771-1781.
[21] 谢卫卫.俯卧位时有创桡动脉与足背动脉血压以及脉压变异度的一致性分析[D].郑州:郑州大学,2018.
[22] 严广斌.视觉模拟评分法[J].中华关节外科杂志:电子版,2014,8(2):273.
[23] Kim SH,Lilot M,Sidhu KS,et al. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure:a systematic review and meta analysis [J]. Anesthesiology,2014,120(5):1080-1097.
[24] 成怀文,权安京,张占军.无创实时动脉血压监测系统在老年患者手术麻醉中的应用价值[J].河南医学研究,2020,29(34):6396-6397.
[25] 宋凌菁,李冬英.ICU应用有创血压监测的护理新进展[J].实用临床医学,2012,13(5):137-139.
[26] 樊丽平.有创动脉血压监测在机械通气病人中的应用及护理[J].全科护理,2014(32):3015-3016. |
|
|
|