Performance evaluation of magnetic bead nucleic acid extraction in real-time fluorescence quantitative PCR detection system for plasma EBV-DNA |
ZHENG Yuhong1 ZHONG Fengjin1 CHEN Yansong1 LIN Yingying1 XIAO Zhenzhou1 PAN Jianji2 LIN Shaojun2 GUO Qiaojuan2 CHEN Yan1▲ |
1.Department of Clinical Laboratory, Fujian Cancer Hospital & Fujian Medical University Cancer Hospital, Fujian Province, Fuzhou 350014, China;
2.Department of Radiotherapy, Fujian Cancer Hospital & Fujian Medical University Cancer Hospital, Fujian Province, Fuzhou 350014, China |
|
|
Abstract Objective To evaluate the application performance of magnetic bead nucleic acid extraction method in real-time fluorescence quantitative PCR detection for plasma Epstein-Barr virus (EBV) DNA. Methods Automated magnetic bead method was used to extract nucleic acid, and combined real-time fluorescence quantitative PCR assays for quantitative EBV-DNA. The detection performance of the method was evaluated by using the third-party reference materials. In addition, 100 patients with NPC were selected, and the plasma EBV-DNA was quantitatively detected by magnetic beads method and boiling method. The correlation between the two methods was analyzed. Results In the automated magnetic bead method-based real-time fluorescence quantitative PCR detection system for EBV-DNA, the repeatability precision of the low, medium and high were all <3/5 TEa (TEa = target ± 0.4 lg), and the intermediate precision were all < 4/5 Tea. In the range of 5.4×101 - 5.4×105 U/mL, there had good linear relationship. The minimum detection limit was about 3.334×101 U/mL. And the concentration of Hb interference in plasma was lower than 8.667 g/L, the results of magnetic bead test were not significantly affected. The detection results between magnetic bead method and boiling method were linearly related, and the qualitative coincidence rate was 83%. Conclusion The detection of plasma EBV-DNA by automated magnetic bead nucleic acid extraction method can achieve good detection performance.
|
|
|
|
|
Cite this article: |
ZHENG Yuhong1 ZHONG Fengjin1 CHEN Yansong1 LIN Yingying1 XIAO Zhenzhou1 PAN Jianji2 LIN Shaojun2 GUO Qiaojuan2 CHEN Yan1▲. Performance evaluation of magnetic bead nucleic acid extraction in real-time fluorescence quantitative PCR detection system for plasma EBV-DNA[J]. 中国医药导报, 2017, 14(24): 13-16,24.
|
|
|
|
URL: |
http://www.yiyaodaobao.com.cn/EN/ OR http://www.yiyaodaobao.com.cn/EN/Y2017/V14/I24/13 |
[1] Le QT,Zhang Q,Cao H,et al. An international collaboration to harmonize the quantitative plasma Epstein-Barr virus DNA assay for future biomarker-guided trials in nasopharyngeal carcinoma [J]. Clin Cancer Res,2013,19(8):2208-2215.
[2] 中国合格评定国家认可委员会.CNAS-CL36:2012医学实验室质量和能力认可准则在分子诊断领域的应用说明[S]. 2015.
[3] NCCLS. CLSI EP7-A2:Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition [S]. 2005.
[4] NCCLS. CLSI EP17-A2:Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures;Approved Guideline-Second Edition [S]. 2012.
[5] NCCLS. CLSI EP6-A:Evaluation of the Linearity of Quantitative Measurement Procedures:A Statistical Approach;Approved Guideline [S]. 2003.
[6] Allen TC. Quality:walk the walk [J]. Arch Pathol Lab Med,2011,135(11):1384-1386.
[7] Allen TC,Hammond ME,Robboy SJ. Quality and the college of american pathologists [J]. Arch Pathol Lab Med,2011, 135(11):1441.
[8] Jennings L,Van DV,Gulley ML. Recommended principles and practices for validating clinical molecular pathology tests [J]. Arch Pathol Lab Med,2009,133(5):743-755.
[9] Guzel O,Guner EI. ISO 15189 accreditation:Requirements for quality and competence of medical laboratories,experience of a laboratory Ⅰ [J]. Clin Biochem,2009,42(4-5):274-278.
[10] Yanikkaya-Demirel G. ISO 15189 accreditation:Requirements for quality and competence of medical laboratories,experience of a laboratory Ⅱ [J]. Clin Biochem,2009,42(4-5):279-283.
[11] 胡荣盛,薛静俊,朱振坤,等.自动核酸提取实时荧光定量PCR检测全血及血浆中EBV DNA载量的性能评价[J].现代实用医学,2013,25(8):940-943.
[12] Thatcher SA. DNA/RNA preparation for molecular detection [J]. Clin Chem,2015,61(1):89-99.
[13] 刘佳,徐军,王雪飞,等.三种HBV DNA提取方法对荧光定量PCR检测结果影响的比较[J].中华检验医学杂志,2008,31(7):780-783.
[14] 陆佳飞,周科隆,王缦.磁珠法快速提取乙型肝炎病毒DNA方法的建立及其应用研究[J].中华检验医学杂志,2012,35(9):843-850.
[15] Hayden RT,Yan X,Wick MT,et al. Factors contributing to variability of quantitative viral PCR results in proficiency testing samples:a multivariate analysis [J]. J Clin Microbiol,2012,50(2):337-345.
[16] 李金明.临床实验室分子诊断的标准化[J].中华检验医学杂志,2006,29(6):483-486. |
|
|
|